Carina Mobile Lifts recall
Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
Why was Carina Mobile Lifts recalled?
A limited amount of Carina mobile lifts have been delivered with a too small locking pin. If a locking pin smaller than the correct size is mounted on the Carina mobile lift, the locking pin is likely to cause increased wear on the plastic bushings. In case of continued use, these components may cause wear on the metal of the piston rod eye.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
a) Carina350EM, REF 6060011, Serial numbers: 661102000 - 661102160 b) Carina350EML, REF 60600012, Serial numbers: 661200500 - 661200584 c) Carina350EE, REF 60600009, Serial numbers: 660900500 -660900571
- Product
- Carina Mobile Lifts, Models: a) Carina350EM, REF 6060011 b) Carina350EML, REF 60600012 c) Carina350EE, REF 60600009
- Recalling firm
- Handicare AB
- Quantity
- 9 units
- Distribution
- OH, PA, MD
- Recall date
- October 18, 2017
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)