Ingenuity CT recall
Ingenuity CT Model Number 728326 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer…
Why was Ingenuity CT recalled?
The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
30003, 30004, 30005, 30006, 30007, 30008, 30009, 30010, 30011, 30012, 30013, 30014, 30015, 30017, 30018, 30019, 30020, 30021, 30022, 30025, 30026, 30027, 30028, 30029, 30030, 30031, 30032, 30033, 30034, 30036, 30037, 30038, 30039, 30040, 30041, 30042, 30043, 30044, 30045, 30046, 30047, 30049, 32045, 32083, 32107, 52051, 52060, 52062, 300003, 300004, 300005, 300009, 300010, 300011, 300012, 300013, 300014, 300015, 300016, 300017, 300018, 300019, 300020, 300021, 300022, 300024, 300025, 300027, 300028, 300029, 300030, 300031, 300032, 300033, 300034, 300035, 300036, 300037, 300038, 300039, 300040, 300041, 300042, 300043, 300044, 300045, 300047, 300049, 300050, 300051, 300052, 300053, 300054, 300055, 300056, 300057, 300058, 300059, 300060, 300061, 300062, 300063, 300064, 300065, 300066, 300068, 300069, 300070, 300071, 300072, 300073, 300074, 300075, 300077, 300078, 300079, 300080, 300081, 300082, 300083, 300084, 300086, 300087, 300088, 300089, 300090, 300091, 300092, 300093, 300094, 300095, 300096, 300097, 300098, 300099, 300100, 300101, 300102, 300103, 300104, 300105, 300106, 300107, 300108, 300109, 300110, 300111, 300112, 300113, 300114, 300115, 300116, 300117, 300118, 300119, 300120, 300121, 300122, 300123, 300124, 300125, 300126, 300127, 300128, 300129, 300130, 300131, 300132, 300133, 300135, 300136, 300137, 300138, 300139, 300140, 300141, 300142, 300143, 300144, 300145, 300147, 300148, 300149, 300150, 300151, 300156, 300157, 300158, 300159, 300160, 300161, 300162, 300163, 300164, 300165, 300166, 300200, 300201, 300202, 300203, 300204, 300205, 300206, 300207, 300208, 300209, 300210, 300211, 300212, 300213, 300214, 300146, 600023, 600047, 300215, 300216, 300217, 300218, 300219, 300220, 300221, 300223, 310058, 310144, 320205, 333057, 333084, 333128, 336061, 336084, 336121, 336233, 30004-1
- Product
- Ingenuity CT Model Number 728326 Product Usage: Computed Tomography X-ray systems intended to product cross-sectional images of the body by computer…
- Recalling firm
- Philips Medical Systems
- Quantity
- 235
- Distribution
- Worldwide Distribution - US Nationwide in the states of:: AK, AL, AR, AZ, CA, CO, CT, DE, GA, IA, IL, IN, FL, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NJ, NH, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, & WY. and the countries of Algeria, Argentina, Australia, Bangladesh, Belgium, Bolivia, Brazil, Cambodia, China, COlumbia, Cuba, Denmark, Dijbouti, Dominican REpublic, Egypt, Estonia, Finland, France, Germany, Gilbraltar, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Latvia, Lebanon, Libya, Lithuania, Malaysia, Martinique, Monaco, Mongolia, Mozambique, Myanmar, Netherlands, New Zealand, Norway, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Seitzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uniter Arab Emirates, United Kingdom, Uzbekistan, and Vietnam.
- Recall date
- October 12, 2016
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (device)