FDAClass IIDecember 12, 2012

Nutropin AQ NuSpin 20 recall

Nutropin AQ NuSpin 20 (somatropin (rDNA origin) injection} 20 mg/ 2mL Genentech, South San Francisco, CA 94080, NDC#50242-0076-01

Why was Nutropin AQ NuSpin 20 recalled?

Miscalibrated and/or Defective Delivery System: Genentech has received complaints for Nutropin AQ NuSpin 10 & 20 reporting that the dose knob spun slowly and the injection took longer than usual (slow dose delivery).

What should I do if I have this product?

Review the official FDA enforcement record for current instructions.

What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.

Is my product affected?

Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.

Which lot codes, UPCs or models are affected?

Lot# 936674 Exp. 09/30/13

Product
Nutropin AQ NuSpin 20 (somatropin (rDNA origin) injection} 20 mg/ 2mL Genentech, South San Francisco, CA 94080, NDC#50242-0076-01
Recalling firm
Genentech Inc
Quantity
10,306 syringes
Distribution
Nationwide
Recall date
December 12, 2012
Classification
Class II
Status
Terminated
Agency
FDA (drug)
Open FDA enforcement record Latest recalls