Curaplex Epi Kit NOT FOR IV USE recall: Labeling Error
Curaplex Epi Kit NOT FOR IV USE, #8600-01120, Epi-Safe Kit 8600-01 120 contains:2 Safety-LOK Syringe, TB, 1cc w/ needle, 25 ga x5/8 , 2 Vials of epinephrine, 1mg/1mL, 4 Alcohol prep pads, 2 Adhesive bandages, 1 x 3 1 Epi Generic Injection. Rx Only. Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016
Why was Curaplex Epi Kit NOT FOR IV USE recalled?
Labeling Incorrect Instructions: This recall has been initiated because the Instructions for Use included in the Epi-Safe Kits for the Epi-Safe Syringe (and/or the Safety Lok Syringe in Model 8600-01120) recommend a midpoint dosage of epinephrine for children between 0 lbs. and 66 lbs. of 0.15mL that has not been approved by FDA for children. Reliance on the current IFU and/or use of the Epi-Safe Syringe as instructed may result in the administration of an inappropriate dosage of epinephrine to young children experiencing anaphylaxis, especially those under 33 lbs. Additionally the Epi Safe syringe does not have the correct markings to allow a health care provider to draw up the correct dosage for children under 33 kg.
What should I do if I have this product?
Review the official FDA enforcement record for current instructions.
What to do with a recalled product: refunds, receipts, safe disposal and how to report a problem.
Is my product affected?
Compare the item in your hand with every identifier below from the official FDA record. Only products that match the listed product and lot, UPC or model codes are part of this recall.
Which lot codes, UPCs or models are affected?
all lot numbers
- Product
- Curaplex Epi Kit NOT FOR IV USE, #8600-01120, Epi-Safe Kit 8600-01 120 contains:2 Safety-LOK Syringe, TB, 1cc w/ needle, 25 ga x5/8 , 2 Vials of epinephrine, 1mg/1mL, 4 Alcohol prep pads, 2 Adhesive bandages, 1 x 3 1 Epi Generic Injection. Rx Only. Distributed by Sarnova HC, LLC s family of companies: Bound Tree Medical, LLC, Cardio Partners, Inc., Emergency Medical Products, Inc. & Tri-anim Health Services, Inc. 5000 Tuttle Crossing Blvd. Dublin, OH 43016
- Recalling firm
- Bound Tree Medical, LLC
- Quantity
- 339 kits (1 standard syringe / 1 IFU per kit)
- Distribution
- Nationwide in the USA
- Recall date
- January 30, 2019
- Classification
- Class II
- Status
- Terminated
- Agency
- FDA (drug)